PerkinElmer / Excelitas PS300-11E140VA Xenon Lamp Ballast Repair
|
We are able to offer a comprehensive repair and refurbishment for the Perkin Elmer / Excelitas PS300-11E140VA series of short arc xenon lamp ballasts. Commonly used within various light sources including the Sutter DG-4 and DG-5 series as well as others manufactured by Stryker, Smith and Nephew, and Applied Biosystems. 
Full functional testing of this product includes:
- Lamp current check / calibration
- HT trigger / ignition voltage pulse check / calibration
- Input line regulation
- Output load regulation
- Output ripple and noise
- Dynamic load testing (regulation response time)
- Interlock function check
- Lamp monitoring signals
- Fan voltage check
- Safety tests (HI-POT / Earth continuity and leakage)
Variants of the PS300-11 serviced to date include :
P300B, PS300-11, PS300-11A, PS300-11-LS, PS300-11A210VA, PS300-11B120VA, PS300-11D225VA, PS300-11E140VA, Y1709, Y1740, Y1745, Y1771, Y1772, Y1773, Y1810, Y1813, Y1822, 226227, 226341-003, 226341-024, 227208, 228032, 228305, 228798, 225050.
Products are fully tested to the original test specification using a combination of electronic loads and a Perkin Elmer 300W short arc lamp.
Depending upon the revision we also upgrade this product to the very latest manufacturing standard.
Shipped with a full 12 months warranty (terms apply).
Fast 24hr turnaround service available (subject to availability of components).
For current service pricing and turnaround times please contact us.